hy-optima1700

H2scan’s HY-OPTIMA 1700 in-line, real time hydrogen-specific process analyzers are designed for ease of use, interface flexibility and true process control. The HY-OPTIMA 1700 is an intrinsically safe device that incorporates our patented solid-state sensor and is configured to operate in background gas streams with up to 95% relative humidity and temperatures up to 60ºC. The HY-OPTIMA 1700 is ideal for process gas streams containing virtually no contaminants and where real time continuous measurements of hydrogen can increase process plant efficiencies, improve diagnostics and enhance maintenance management translating into higher profitability.

 




Performance

Hydrogen Sensitivity Range:
0.5% to 100% hydrogen by volume at 1 ATM

Response Time: T90 < 30 sec
Ingress Protection: IP64 capable
Recommended Verification Interval: 90 days
Product Life Expectancy: 10 years
Accuracy(*):
± 0.3% absolute for 0.5 to 10% H2
± 1.0% absolute for 10 to 100% H2
Drift/week:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Repeatability:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Linearity:
± 0.2% absolute for 0.5-10% H2
± 0.4% absolute for 10-100% H2

Interface Options

Power Barrier Input Voltage: 24 VDC nominal
Power Barrier Input Voltage Range: 20 VDC to 28 VDC
Intrinsically Safe Sensor Input Voltage: 10 VDC
Sensor Input Voltage Range: 5 VDC to 28 VDC
Input Power: 10 W
Analog Output: 4 mA to 20 mA
Serial Communication: RS422

Certifications

CE 0359, Ex
II 1 G, Ex ia II H2, d+d IIB T3
II 2 G, Ex ib II H2, ib d IIB T3
ITS07ATEX25634X

Operating Conditions

Operating Humidity: 0% to 95% RH
Flow Rate: 0.1 slpm to 10 slpm
Pressure: 0 to 2 bar gauge, 0 to 30 psi gauge
Process Gas Temperature Range: -20ºC to 60ºC
Operating Temperature Range: -20ºC to 40ºC
Storage Temperature Range: -40ºC to 50ºC
Calibration Background Gas: Nitrogen

Included Accessories

Intrinsically Safe Power Barrier
Armored Power / Analog Cable: 4 m
4m Serial Cable
IS Power Barrier
Fitting Adapter
Calibration Cup

Optional Accessories

Single Channel Intrinsically Safe Analog Output Barrier (with or without relays)

Single Channel Intrinsically Safe RS422 Serial Interface Barrier

Serial Interface Cable: 4 m (custom lengths available)
Armored Power / Analog Cable (custom lengths available)

Adapter Fittings

½ in. MNPT thread
½ in. FNPT thread
-8 SAE/MS thread size
... and many other industry standards

Dimensions

Total Length: 10.3 in.
Length of Enclosure: 6.8 in.
Length of Tube: 2.9 in.
Width: 3.4 in.
Depth: 1.4 in.

 

Download specifications